One Step Fast RT-qPCR Probe Premix-UNG
Cat No: HCR5143A
Unus Gradus RT-qPCR Probe Kit (FOR FAST) est probe-substructio RT-qPCR ieiunium detectionis ornamentum aptum ad singula plex vel multiplex quantitatis PCR utens RNA in template (ut RNA virus).Productum hoc utitur nova generatione antibodyi mutato Taq DNA Polymerase et uni gradatim dicato Reverse Transcriptase, cum quiddam optimized ad celeri amplificationem, quae celerius amplificationis celeritatem, amplificationem efficientiam et proprietatem altiorem habet.Aequatum amplificationem sustinet in utroque uno-plexo et multiplex gravis et altae in brevi concentratione.
Components
1. 5×RT-qPCR quiddam (U+)
2. Enzyme mix (U+)
Notae:
a.5×RT-qPCR quiddam (U+) includit dNTP et Mg2+;
b.Enzyme mix (U+) includit transpositionem transcriptase, Hot Committitur Taq DNA polymerase, RNase inhibitor et UDG;
c.Utere RNase-Free apices, EP tubes, etc.
Ante usum, 5×RT-qPCR quiddam permisceto (U+).Si quid post tabidaque praecipitatio est, exspecta quiddam ad reditum ad cella temperiem, misce et dissolve, et tunc utere sole.
Repono Conditions
Productum siccis glaciem evehitur et in -25~-15℃ pro 1 anno condi potest.
Instructiones
1. reactionem Systema
Components | Volumen (20 μL reactionem) |
2 RT-qPCR quiddam | 4μL |
Enzyme misce (U+) | 0.8μL |
De primario | 0.1~ 1.0μM |
Primario Reverse | 0.1~ 1.0μM |
TaqMan Probe | 0.05~0.25μM |
Formula | X μL |
RNase-Free aquae | ad 25μL |
Notae: Reactio volumen 10-50μL est.
2. Protocollum revolutio (Standard)
Cycle step | Temp. | Tempus | Cycles |
Reverse Transcription | 55 | 10 min | 1 |
Denaturation Coepi | 95 | 30 sec | 1 |
Denaturation | 95 | 10 sec | 45 |
Annealing / Extensio | 60 | 30 sec |
Revolutio Protocol (Celer) Cycle step |
Temp. |
Tempus |
Cycles |
Reverse Transcription | 55 | 5 min | 1 |
Denaturation Coepi | 95 | 5 s | 1 |
Denaturation | 95 | 3 s | 43 |
Annealing / Extensio | 60 | 10 s |
Notae:
a.Inversa transcriptio temperatura est inter 50℃ ad 60℃, temperatura augens adiuvat ad amplificandas structuras multiplices et altas CG contentas templates;
b.Optima temperatura furnum debet aptari secundum Tim valoris primitivae, et eligere brevissimum tempus pro fluorescentiae signo collectionis innixa in Vero tempore PCR instrumenti.
Notae
Quaeso uti necessariam PPE, talem lab tunica et caestus, ut tuam salutem et salutem curet!